Kai Chao Brand BD Warning Test Kit
Price:¥
【Product Name】Kaizhao Brand B-D Warning Test Kit
【Executive Standard】Q/AHKC019-2023
【Specification Model】KC2159-5
【Color Change Description】After sterilization, if the chemical indicator on the label of the test kit changes from light yellow to dark brown or black, it indicates that the test kit has been exposed to saturated steam.
Qualified: The pattern on the test chart changes from the original light yellow to a uniform dark brown or black, i.e., the color of the central part and the edge part is consistent, the B-D test is qualified, indicating that the air discharge is thorough, the sterilizer works well without leakage, and can be used normally.
Unqualified: The pattern on the test chart does not change color or does not change color thoroughly, usually the central part is lighter in color than the edge part, the B-D test is unqualified, indicating that the air discharge is not thorough or there is a leakage, the sterilizer is faulty, and it must be checked and repaired before it can be used. Early Warning: If the light yellow circular indicator in the center of the warning test chart changes color lightly than the standard color or does not change color, it indicates that the sterilizer is approaching a fault and needs to be inspected as soon as possible.
【Scope of Use】Suitable for the air discharge and steam permeability test of 134℃, 3.5min pre-vacuum pressure steam sterilizer.
【Usage Method】
This product does not need to be used with the standard test kit specified in the "Disinfection Technology Specification". When using, you can directly place the test kit at the exhaust port of the sterilizer, close the cabinet door, and carry out a 134℃, 3.5-minute sterilization test treatment. When finished, open the cabinet door, open the test kit, take out the test chart, and observe the test results.
【Precautions】
1. The test chart is prohibited from contacting acid, alkali, and must not be damp (relative humidity should be less than 50%) when stored and used.
2. Store in a dark place, away from ultraviolet rays, fluorescent lights, and sunlight.
3. The test is carried out under 134℃ steam conditions, and the time should not exceed 4 minutes.
4. The test is performed in an empty锅 before the daily***sterilization.
5. This product cannot be used as a means to detect the effectiveness of steam sterilization.
【Validity Period】The validity period is 12 months. The packaging specifications, production date, and batch number are indicated on the product packaging.
Online inquiry